Sistema de selección en línea
Bienvenido a nuestra biblioteca completa de instrumentos y sensores. Encuentre rápidamente el producto y la solución adecuados para su aplicación. Comience con una búsqueda o elija un producto.

Technical articles are currently available in Chinese and English. English content is shown below.

← Back to knowledge hub
Selection and applications4 min readUpdated 2026-10-14

How to read accuracy, turndown and zero stability

A headline percentage may be insufficient at low flow. Compare definitions, operating ranges and test conditions.

Start an application checklist

Check what the percentage refers to

Percent of reading and percent of full scale use different reference quantities and cannot be compared directly. Look for additional fixed terms, minimum-flow conditions and fluid restrictions. Accuracy, repeatability, resolution and system uncertainty also describe different things.

Why zero stability matters at low flow

Coriolis data sheets often express zero stability in mass-flow units. At low flow, this term can become a larger proportion of the measured flow. Some manufacturers specify a low-flow calculation, but it must not be transferred to another model without verification. Turndown alone does not establish error throughout the operating range.

Compare two operating points

For illustration only, a 0.5 kg/h error term represents 0.5% of 100 kg/h and 5% of 10 kg/h. This is a ratio calculation, not a meter performance claim, and it does not replace the manufacturer's complete error model. Differences between calibration and field conditions also require assessment.

  • Request performance at minimum, normal and maximum flow.
  • Distinguish meter performance, installation effects and reference-instrument uncertainty.

Ask for evidence that supports comparison

Request performance definitions, applicable flow ranges, zero stability, fluid and temperature/pressure conditions, and calibration or verification information. For batching or commercial measurement, review the complete measurement chain and totalisation requirements. A display reading zero once is not evidence that zero stability meets a specification.

References and further reading

Coriolis application checklist

Gather fluid, flow, pressure/temperature and installation conditions to identify missing information and engineering checks.

Start an application checklist
¿Aún no ha iniciado sesión en DOWESTON?
Cree una cuenta para empezar a seleccionar, consultar precios y recibir nuestro boletín con las últimas novedades.
Productos y servicios
Doweston ofrece una gama de servicios para ayudarle a construir rápidamente sistemas de instrumentación automatizada. Contáctenos para más información.
¿Listo para comprar?
Si desea comprar productos,
solo contacte con nuestro equipo de ventas — estaremos encantados de ayudarle.
Quiénes somos
DOWESTON ofrece soluciones de medición precisas y fiables
¡productos de detección y sensores fiables para todos los sectores!
¿Necesita servicio?
Si desea optimizar la producción y reducir los parones,
podemos ayudarle a elegir las piezas de repuesto y los productos de servicio adecuados.
Editar configuración

Producto

Marca

Modelo

Confirmar cambios Cancelar
Su pantalla puede ser demasiado pequeña para una mejor experiencia.
Le recomendamos continuar en un ordenador,
¡o contacte de inmediato con un ingeniero para su selección!
¿Servicio más rápido? Llame al
400-888-4436
Seguir navegando