Field troubleshooting3 min readUpdated 2026-10-14

Why is flow displayed when the process appears stopped?

Check the actual no-flow condition, filling, stability, display variable and status before treating zero calibration as the solution.

Start an application checklist

Identify what the display represents

Record model, alarms, units, displayed variable and the reading over time. Instantaneous flow, a totaliser and a control-system calculation are different values. A totaliser retaining its historical value does not mean flow continues. Record positive or negative readings and timing rather than sending only one screenshot.

A closed valve is not proof of zero flow

Use existing operating records and process information to review bypasses, recirculation and possible valve leakage. Do not operate valves independently for this check. For liquid service, site personnel must also establish whether the tube is full, gas or phase changes are present, and temperature and pressure are stable.

Look for process and configuration clues

Observe whether the symptom coincides with pump operation, nearby vibration or process changes. Review direction, units, damping and low-flow cutoff through existing screens or configuration records. A cutoff may affect small displayed values; it does not establish that the underlying cause has been removed.

  • Record flow, density, temperature and alarms during the apparent no-flow period.
  • Keep existing configuration and recent change records for comparison.

When to involve an engineer

Refer persistent offset, alarms or uncertain no-flow/filling conditions to an engineer, who can determine whether zero verification or calibration is appropriate under the model's manual. This page is for observation and document checks, not field zeroing instructions. Opening covers, wiring, pipe removal, pressurised work and hazardous-area tasks require site procedures.

References and further reading

Coriolis application checklist

Gather fluid, flow, pressure/temperature and installation conditions to identify missing information and engineering checks.

Start an application checklist
¿Aún no ha iniciado sesión en DOWESTON?
Cree una cuenta para empezar a seleccionar, consultar precios y recibir nuestro boletín con las últimas novedades.
Productos y servicios
Doweston ofrece una gama de servicios para ayudarle a construir rápidamente sistemas de instrumentación automatizada. Contáctenos para más información.
¿Listo para comprar?
Si desea comprar productos,
solo contacte con nuestro equipo de ventas — estaremos encantados de ayudarle.
Quiénes somos
DOWESTON ofrece soluciones de medición precisas y fiables
¡productos de detección y sensores fiables para todos los sectores!
¿Necesita servicio?
Si desea optimizar la producción y reducir los parones,
podemos ayudarle a elegir las piezas de repuesto y los productos de servicio adecuados.
Editar configuración

Producto

Marca

Modelo

Confirmar cambios Cancelar
Su pantalla puede ser demasiado pequeña para una mejor experiencia.
Le recomendamos continuar en un ordenador,
¡o contacte de inmediato con un ingeniero para su selección!
¿Servicio más rápido? Llame al
400-888-4436
Seguir navegando