Electromagnetic flowmeters: how to check the minimum liquid conductivity

Use conductivity at the actual process temperature and concentration to check the model requirement. The name of an aqueous solution alone is insufficient to establish measurability.

Commencer une fiche d’application

Check the operating range

Check the model against the liquid conductivity at actual operating temperatures and concentrations. Identifying the fluid as an aqueous solution does not establish that the meter can measure it.

Include the lowest concentration and the temperature extremes. A single laboratory result for the formulation at room temperature cannot represent the full process range.

What to verify and when it applies

The conductive liquid, the electrode interface and the reference connection form the path for a weak signal. Check that the liquid maintains effective electrical contact, then look for electrode fouling or changes in the reference connection. A reading during one process stage shows that there is a signal at that stage. It does not establish an equal signal margin for every stage.

Obtain both the conductivity of the actual process liquid and the minimum conductivity specified for the flowmeter. To establish interchangeability, compare mechanical connections, electrical connections, functions and calibration compatibility. Similar appearance or connections cover only part of the comparison. Check the software options and accessories against the supply records. When the configuration is finalized, retain the full model code, drawings and verification requirements. After servicing, confirm which elements of the original configuration the replacement affected.

How to check the application

  • Obtain the conductivity of the actual liquid and the meter’s minimum requirement. Record available flow readings and electrode or empty-pipe diagnostics. Identify which data were actually obtained and which functions are documented for this meter.
  • List the minimum, typical and most unfavorable conductivity conditions, recording their durations and the source of each value.
  • Compare before and after conditions using a reference suitable for the instrument and the application. Keep the original conductivity-related settings and process conditions on record. Address one factor at a time based on the evidence.

Example: unstable readings as conductivity drops

Suppose a pure-water process gives an unstable signal as conductivity falls. If the liquid at its lowest concentration falls below the model requirement, stable readings at a higher concentration cannot establish suitability for the entire formulation cycle.

First compare the current process conditions with those used to select the original meter.

Record the result and the delivered configuration

State the actual conductivity results beside the meter’s minimum requirement. Record which operating conditions were covered, the basis for the comparison and any open items. Keep the original measurements and configuration with the report. A test that simulates only the output cannot verify the parts of the measurement chain it bypasses.

This is an illustrative case. Follow the actual model limits and site procedures. Confirm the supplied specification and certification from the Doweston product documentation.

Appliquez ces connaissances à la sélection

Clients et achats

Vérifiez les conditions réelles et les critères de réception dans l’article et la documentation du produit. Soumettez les points ouverts au service technique DOWESTON.

Équipes commerciales et techniques

Notez les questions du client et ses conditions de service. Vérifiez la plage, les raccordements et les conditions d’utilisation dans la documentation du modèle sélectionné.

Transformez les questions en critères de sélection

Relevez le fluide, les raccordements, le montage et les critères de réception du produit, puis examinez les points ouverts avec DOWESTON.

Commencer une fiche d’application
Pas encore connecté à DOWESTON ?
Créez un compte pour commencer la sélection, consulter les prix et recevoir notre lettre d'information sur les dernières nouveautés.
Produits et services
Doweston propose une gamme de services pour vous aider à construire rapidement vos systèmes d'instrumentation automatisés. Contactez-nous pour en savoir plus.
Prêt à acheter ?
Si vous souhaitez acheter des produits,
contactez simplement notre équipe commerciale — nous serons ravis de vous aider.
Qui sommes-nous
DOWESTON fournit des solutions de mesure précises et fiables
— des produits de détection et de capteurs fiables dans tous les domaines !
Une demande de service ?
Si vous souhaitez optimiser votre production et réduire les arrêts,
nous pouvons vous aider à choisir les pièces de rechange et les services adaptés.
Modifier la configuration

Nom du produit

Marque

Modèle

Confirmer les modifications Annuler
Votre écran est peut-être trop petit pour une navigation optimale.
Nous vous recommandons de continuer sur un ordinateur,
ou contactez immédiatement un ingénieur pour votre sélection !
Service plus rapide ? Appelez le
400-888-4436
Continuer la navigation