La version française n’est pas encore disponible. La version complète en anglais est présentée ci-dessous.
← Retour à la base de connaissancesMultiaxis force and torque transducers: 50 selection, application and maintenance questions
50 shared questions for Multiaxis force and torque transducers, covering selection, delivery, maintenance, common questions and evidence-based brand comparisons, with verification actions, limits and sources.
Conservez vos ressources techniques
Conservez un brouillon dans ce navigateur, puis importez-le manuellement dans votre bibliothèque après connexion.
La création de compte suit le processus habituel d’inscription et d’approbation. L’enregistrement de ressources n’ouvre pas automatiquement l’accès aux prix ni aux autres autorisations commerciales.
Shared guidance for this product family. Confirm the complete model and application; family-level guidance is not evidence of a particular product specification or approval.
Les principes généraux s’appliquent en complément de la validation du modèle. Les performances, homologations et alternatives dépendent du code de commande complet et des conditions réelles.
1What does the Multiaxis force and torque transducers principle establish, and what does it not establish?
A multiaxis transducer measures force and moment components in a defined coordinate system, often using multiple strain bridges and calibration relationships. Axis ratings, crosstalk and combined loads require separate review. The principle explains signal generation; it does not establish the range, accuracy, materials or approvals of every model. Review the physical principle and the application limits separately.
Verification: Confirm "Required force/moment components and positive coordinate directions" and assess "Individual axis ratings do not prove every combined load is permissible" against the actual application. Record the complete model and document revision used.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
2Is the required result measured directly or calculated from other inputs?
Define the quantity to be accepted before selecting a display unit. A multiaxis transducer measures force and moment components in a defined coordinate system, often using multiple strain bridges and calibration relationships. Axis ratings, crosstalk and combined loads require separate review. For "Required force/moment components and positive coordinate directions", document output definitions and conversion or compensation conditions; displaying a value does not make it a direct measurement under all conditions.
Verification: Document the process objective, units and reference conditions. Check "Combining all axis maxima without checking the load envelope" and list separate inputs and acceptance methods for calculated results.
Limites d’application : Combining all axis maxima without checking the load envelope
3Which application information is needed before requesting a quotation?
Prepare at least these five input groups: Required force/moment components and positive coordinate directions; Working, peak and simultaneous combined loads on each axis; Load point, line of action and reference origin; Sampling synchronisation, bandwidth and decoupling requirements; Mounting flanges, preload, temperature and cable environment. Mark unknown items for confirmation instead of substituting a normal point or an assumed default for all conditions.
Verification: Have process, instrumentation and purchasing roles agree the inputs, particularly "Working, peak and simultaneous combined loads on each axis". Resolve missing evidence before requesting application confirmation.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
4How should the range be chosen rather than copied from a nominal equipment rating?
Select the range from actual operating conditions and permissible error, then check extremes and installation constraints. For this family: Individual axis ratings do not prove every combined load is permissible Nominal size or maximum equipment rating need not equal the useful measuring interval.
Verification: List minimum, normal, maximum and abnormal conditions. Check "Load point, line of action and reference origin" against the documented performance envelope.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
5Can a larger turndown claim or display rescaling fix poor low-end performance?
No. Low-end usability depends on the principle, available signal and error mechanisms. Check crosstalk definitions and the affected-axis full-scale reference More display digits, software scaling or a stated turndown cannot remove the original noise or installation error.
Verification: Obtain error definitions and limits at the actual low operating point. Use "Scale/sign error: inspect units, coordinate transform and sensitivity" to investigate bias instead of comparing only percent-of-full-scale values.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
6How should compatibility with the measured object and materials be established?
Use the complete measured object and actual exposure, rather than a material name alone. Sampling synchronisation, bandwidth and decoupling requirements Avoid this family-specific misuse: Treating force rotation as a complete moment transformation For non-wetted equipment, assess loads, target materials, signals and environmental compatibility instead.
Verification: List the actual contact or loading parts, seals, insulation and cleaning conditions. Have the responsible parties assess "Moments depend on reference origin; coordinate or unit errors alter results".
Limites d’application : Moments depend on reference origin; coordinate or unit errors alter results
7Why must abnormal and environmental conditions be listed separately from normal operation?
Normal measurement, temporary excursions and survival limits are different requirements. Moments depend on reference origin; coordinate or unit errors alter results Start-up, cleaning, impacts, power interruption or abnormal environments can determine the necessary construction and protection.
Verification: Complete the worst-case conditions using "Mounting flanges, preload, temperature and cable environment" and check "A compensation matrix must match the sensor and calibrated scope". Separate continuous operating limits from temporary limits in the specification.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
8Can the published instrument accuracy be used as the installed system accuracy?
Not without assessment. Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance. The installed system also includes object, mounting, signal conversion, environmental and reference effects. Derived results additionally depend on the relevant input uncertainties.
Verification: Agree actual test points, installation conditions and the acceptance reference. Check "Mount by locating surfaces and axis marks with controlled preload and moments" and document the system error or uncertainty basis.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
9How should response time, filtering and a stable display be balanced?
A stable indication and fast change detection involve a trade-off; filtering cannot repair incorrect installation or real process variation. Multiple changing axes: distinguish eccentric loading from crosstalk Review the complete chain when fast response matters, not merely the display refresh rate.
Verification: Define the shortest relevant event and acceptable delay. Check "Establish baseline and tool weight and verify signs with known load directions" and verify local output, display and receiving-system response using a controlled change.
Limites d’application : Retaining a matrix after channel or unit changes
10Which typical applications fit, and which require a specific engineering review?
Typical contexts include: Robot tool contact-force monitoring; Multidirectional load test stands; Assembly force and moment verification. Similar application labels do not establish equivalent conditions. In particular, assess "A compensation matrix must match the sensor and calibrated scope" before relying on another installation as proof.
Verification: Select the nearest application and compare "Required force/moment components and positive coordinate directions" and "Mounting flanges, preload, temperature and cable environment". Obtain written confirmation where published application limits are exceeded.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
11How should a complete order code be defined without unspecified standard options?
Translate required functions, performance and physical interfaces into verifiable options. Axis count, directions, reference origin and moment units; Axis ratings and manufacturer combined-load criteria. Options affecting suitability or supply scope should be explicit; terms such as standard or same as existing are insufficient.
Verification: Check the complete order code and selection sheet, especially "Load point, line of action and reference origin". List accessories, documents, calibration and delivery boundaries separately.
Limites d’application : Combining all axis maxima without checking the load envelope
12Which mounting and mechanical layout checks belong before ordering?
Mount by locating surfaces and axis marks with controlled preload and moments The installation drawing must match the available space and process or mechanical structure. Access for fitting and servicing should be established before delivery.
Verification: Review a dimensioned, oriented site drawing. Use "Sampling synchronisation, bandwidth and decoupling requirements" and "Moments depend on reference origin; coordinate or unit errors alter results" to confirm connections, support, removal access and safe isolation.
Limites d’application : Moments depend on reference origin; coordinate or unit errors alter results
13How can supporting infrastructure prevent an instrument that fits but cannot be used?
Use matching bridge conditioning and verify every channel identity Supporting conditions are part of the measurement system. Matching mechanical connections or signal names alone do not establish power, I/O, grounding or configuration compatibility.
Verification: Confirm "Mechanical location, flanges and preload requirements" and "Bridge channels, synchronisation and matrix format" against actual terminals, loops and receiving-system settings.
Limites d’application : Treating force rotation as a complete moment transformation
14What is a sensible commissioning sequence before changing correction factors?
Check identity, installation and basic conditions first, then status, output and process response. Establish baseline and tool weight and verify signs with known load directions Preserve original evidence for unexplained anomalies instead of concealing them with zero or factor adjustments.
Verification: Record factory settings and initial status. Investigate "Missing axis: inspect bridge, mapping and matrix dimensions" step by step and change parameters only after identifying the cause and using an authorised procedure.
Limites d’application : Retaining a matrix after channel or unit changes
15Do identical output units or interface names establish connection compatibility?
No. Establish the represented object and reference conditions: Axis count, directions, reference origin and moment units. Electrical equipment also needs loop type, scaling, failure behaviour and protocol details checked. Mechanical indicators require clear boundaries between indication and remote-transmission accessories.
Verification: Check the receiving-system variable against "Required force/moment components and positive coordinate directions" and verify limits, units and abnormal states with the actual equipment. Record wiring, configuration and end-to-end checks against "Use matching bridge conditioning and verify every channel identity".
Limites d’application : Treating force rotation as a complete moment transformation
16What should a calibration report contain to be relevant to the application?
Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance. The report should identify the delivered instrument, measured quantity, conditions, method and results. A general conformity certificate does not establish performance for every option or operating condition.
Verification: Check serial number, points, units, reference equipment and traceability. Compare "Working, peak and simultaneous combined loads on each axis" with the calibrated conditions and agree separate acceptance where they differ.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
17How can delivery and site acceptance use reproducible pass criteria?
Verify coordinates, crosstalk and processing with known single and combined loads Agree criteria before testing, including range, conditions, error definition and treatment of anomalies. Similar-looking readings or an unsuitable portable reference are not sufficient acceptance evidence.
Verification: Choose an appropriate reference using "Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance.". Record "Required force/moment components and positive coordinate directions" and test conditions; investigate "Scale/sign error: inspect units, coordinate transform and sensitivity" before deciding on rework or recalibration.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
18What belongs in the quotation and agreement for a non-standard application?
Define why the standard supply is insufficient, the required verification and the responsible parties. A compensation matrix must match the sensor and calibrated scope; Ignoring real moments from tool offset and weight. Additional materials, construction, documents or tests depend on actual requirements.
Verification: Attach "Mounting flanges, preload, temperature and cable environment" and the deviations. Obtain technical confirmation, additional charges, delivery conditions and acceptance methods; unknown requirements must not be assumed satisfied.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
19Where should investigation start when there is no valid reading or output?
Missing axis: inspect bridge, mapping and matrix dimensions Distinguish the actual measured condition from device-state and readout or transmission faults. Do not immediately interpret missing data as sensor failure or a zero process value.
Verification: Under appropriate safe isolation, check "Use matching bridge conditioning and verify every channel identity". Compare local and receiving-end status and record when the fault appeared.
Limites d’application : Combining all axis maxima without checking the load envelope
20How can unstable readings be separated from genuine process variation?
Multiple changing axes: distinguish eccentric loading from crosstalk Compare independent process evidence and raw signals before deciding whether installation, interference or equipment is responsible. More filtering can hide a real abnormal event.
Verification: Keep conditions comparable and record correlation with process or equipment events. Check "Mount by locating surfaces and axis marks with controlled preload and moments" and change one factor at a time.
Limites d’application : Retaining a matrix after channel or unit changes
21How should a persistent high or low bias be investigated before changing factors?
Scale/sign error: inspect units, coordinate transform and sensitivity Systematic bias can arise from units, reference conditions, configuration, installation or the reference method. An unexplained correction can move one operating-point error to other conditions.
Verification: Align quantities, time bases and units first. Review the reference using "Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance." and confirm that "Working, peak and simultaneous combined loads on each axis" matches the configuration.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
22When is it appropriate to adjust a drifting zero or reference position?
Zero changes: inspect tool weight, preload, temperature and collision Establish a genuine, stable reference condition that meets the method requirements before considering adjustment. Some instruments have no freely adjustable site zero.
Verification: Check "Establish baseline and tool weight and verify signs with known load directions" and "Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance." and retain before-and-after results. If the conditions cannot be established, remove external effects or use a qualified service.
Limites d’application : Retaining a matrix after channel or unit changes
23Which evidence helps diagnose intermittent alarms or failures?
Combined-load anomaly: inspect load-envelope limits and mounting stiffness Intermittent faults need correlated event records, not just a normal display after recovery. Review alarm settings, communication status and actual process abnormalities separately.
Verification: Retain time, status codes, raw trends and the contemporaneous "Mounting flanges, preload, temperature and cable environment". Compare "Use matching bridge conditioning and verify every channel identity" and "Establish baseline and tool weight and verify signs with known load directions" to locate the failing link.
Limites d’application : Ignoring real moments from tool offset and weight
24Why should repeated failures trigger a cause review instead of another identical replacement?
Repeated failures warrant a common-cause assessment. Ignoring real moments from tool offset and weight If the external conditions remain unchanged, an identical construction or installation can experience the same failure again.
Verification: Summarise failed parts, service time and "Mounting flanges, preload, temperature and cable environment". Check "Mount by locating surfaces and axis marks with controlled preload and moments" and "Moments depend on reference origin; coordinate or unit errors alter results" and establish corrective action before selecting repair or redesign.
Limites d’application : Moments depend on reference origin; coordinate or unit errors alter results
25What distinguishes a site comparison, an adjustment and a calibration?
Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance. A comparison identifies differences; adjustment changes indication or parameters; calibration establishes a reference relationship under stated conditions. They are not interchangeable claims.
Verification: Define what performance and evidence must be established, using conditions relevant to "Working, peak and simultaneous combined loads on each axis". Retain original and adjusted results, references and responsibility records.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
26What must be checked again after replacing an instrument with the same basic model?
The same basic model does not establish identical options, software, calibration or accessories. Bridge channels, synchronisation and matrix format; Calibration, crosstalk and temperature evidence. Previous system operation does not remove the need to verify a replacement.
Verification: Compare full order codes and revisions. Recheck critical points, failure states and interfaces using "Verify coordinates, crosstalk and processing with known single and combined loads" before handover.
Limites d’application : Ignoring real moments from tool offset and weight
27Can every published maximum rating be achieved simultaneously?
No. Individual axis ratings do not prove every combined load is permissible Maximum values usually relate to particular options and conditions. Temperature, load, pressure, materials, speed or signal conditions may constrain one another.
Verification: Review "Load point, line of action and reference origin" and "Mounting flanges, preload, temperature and cable environment" together. Obtain confirmation for the combined conditions instead of assembling unrelated maxima from separate pages.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
28Does a missing signal or zero indication establish that the process value is zero?
No. Missing axis: inspect bridge, mapping and matrix dimensions Indication, measurement and transmission have distinct possible failure modes. Also consider "Combining all axis maxima without checking the load envelope"; establish the actual process condition independently.
Verification: Check available status or failure indications and "Mechanical location, flanges and preload requirements". For interlocks, follow the complete safety design rather than substituting a zero value for diagnosis.
Limites d’application : Combining all axis maxima without checking the load envelope
29Does one acceptable test point establish compliance across the working range?
No. Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance. Test points and directions must reflect the working interval, relevant error mechanisms and acceptance purpose. Multiple points, reversal or load tests depend on the method.
Verification: Define acceptance coverage against "Working, peak and simultaneous combined loads on each axis" and include critical operating points. Mark conditions outside the tests as unverified.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
30Are the largest range, highest accuracy class and most options always preferable?
Not necessarily. Options should serve the required objective; excess configuration can add cost, complexity or incompatibility. Check crosstalk definitions and the affected-axis full-scale reference; Treating force rotation as a complete moment transformation. Establish essential requirements before comparing optional benefits.
Verification: Document the objective in "Required force/moment components and positive coordinate directions" and why "Axis count, directions, reference origin and moment units" is needed. Compare performance at actual working points rather than one headline metric.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
31Can every maintenance or recalibration interval simply be fixed at one year?
Not independently of risk and history. Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance. Intervals depend on applicable law or contract, conditions, drift, maintenance and failure evidence. Generic advice cannot replace mandatory requirements.
Verification: Use "Zero changes: inspect tool weight, preload, temperature and collision" and "Ignoring real moments from tool offset and weight" to establish and review an initial plan. Retain the justification and approval for interval changes.
Limites d’application : Ignoring real moments from tool offset and weight
32Can parts from different brands be mixed if their physical interfaces fit?
Physical fit is insufficient. Axis ratings and manufacturer combined-load criteria; Mechanical location, flanges and preload requirements. Matching, materials, signals, calibration and approval boundaries can change with the combination.
Verification: Check the complete combination against "Use matching bridge conditioning and verify every channel identity" and obtain compatibility and acceptance evidence. An unconfirmed assembly must not be described as an approved system.
Limites d’application : Treating force rotation as a complete moment transformation
33Does a hazardous-area, hygienic or safety label establish suitability for that duty?
Check the exact order configuration and valid documents rather than a series name. Calibration, crosstalk and temperature evidence "A compensation matrix must match the sensor and calibrated scope" also matters; certificates, installation conditions and restrictions together define suitability.
Verification: Obtain certificate identifiers, covered models and special conditions and match them to the site. Similar names do not grant functional-safety or legal-metrology approval.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
34What evidence makes the final result acceptable, reproducible and sustainable?
Establish the chain from quantity definition and application limits through configuration, mounting, reference and operating status. Verify coordinates, crosstalk and processing with known single and combined loads A normal indication or brochure alone cannot demonstrate continuing system reliability.
Verification: Put "Required force/moment components and positive coordinate directions", "Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance." and "Combined-load anomaly: inspect load-envelope limits and mounting stiffness" into the technical, acceptance and maintenance records, with responsibility for reviewing excursions.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
35Can a DOWESTON model replace an installed instrument from another brand?
Possibly, after a full configuration and application comparison. A basic model name or matching output is not proof of equivalent replacement. Axis count, directions, reference origin and moment units; Axis ratings and manufacturer combined-load criteria; Mechanical location, flanges and preload requirements. Record conclusions on material differences.
Verification: Provide the original full order code, data sheet and site conditions. Check "Bridge channels, synchronisation and matrix format" and "Calibration, crosstalk and temperature evidence" and obtain the necessary test and approval evidence before replacement.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
36Why choose DOWESTON, and what would make that reasoning evidence-based?
Base the decision on project compliance and documented delivery or service, rather than a slogan. "Required force/moment components and positive coordinate directions" and "Individual axis ratings do not prove every combined load is permissible" are important family-specific checks. Claim only benefits supported by the offered product documents and written commitments.
Verification: Request the proposed configuration matching "Axis count, directions, reference origin and moment units", test evidence and delivery/service terms. Do not infer unverified performance, approvals, origin or service life.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
37How can competing brands be compared beyond a single accuracy or price figure?
Use identical conditions and supply boundaries. Axis count, directions, reference origin and moment units; Mechanical location, flanges and preload requirements; Calibration, crosstalk and temperature evidence. Quotes with different ranges, compensation, accessories or calibration scope can conceal material differences.
Verification: Compare "Working, peak and simultaneous combined loads on each axis", "Check crosstalk definitions and the affected-axis full-scale reference" and mounting/acceptance conditions. Separate demonstrated, unverified and inapplicable items; do not claim unproven overall superiority.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
38How should purchasing compare cost without overlooking additions or rework?
Compare complete supply and use-cycle costs, not only the main instrument price. Mount by locating surfaces and axis marks with controlled preload and moments; Calibrate forces and moments traceably in defined coordinates and origins, retaining axis sensitivities, crosstalk and applicable matrices; a single-axis check does not establish all combined-load performance. List the actual installation, supporting equipment, calibration, consumables, downtime and service requirements without assuming any brand is always cheaper.
Verification: Use the same "Axis count, directions, reference origin and moment units" configuration and "Mounting flanges, preload, temperature and cable environment" boundary. Separate equipment, accessories, commissioning and maintenance responsibilities and retain a traceable comparison.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
39What are the six components of a six-axis force sensor?
They are normally three force components and three moments, but actual components and signs follow the device definition. Moment is not another force direction and has different units and reference-point dependence.
Verification: Apply known directional loads channel by channel and record signs, units and origin.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
40Why can a sensor overload while every axis is below its individual rating?
Simultaneous forces and moments combine stresses in the elastic structure. Manufacturer combined-load criteria can limit combinations below individual maxima.
Verification: List simultaneous load combinations and assess them with the model’s documented criterion and dynamic peaks.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
41Does output on other axes under one-direction loading mean a fault?
Not necessarily. Crosstalk can occur, or an off-centre load can create a genuine moment. Verify loading geometry before assessing the specified crosstalk.
Verification: Use a well-defined fixture and compare all axis outputs with documented crosstalk limits.
Limites d’application : Moments depend on reference origin; coordinate or unit errors alter results
42Why can moment zero change after changing the tool?
Tool weight and centre of gravity create real forces and moments about the sensor origin; orientation changes gravity components. Define the compensation model before calling it drift.
Verification: Record tool mass, centre of gravity and orientation and check compensation at known poses.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
43How should moments be treated when changing coordinate systems?
If the origin moves, include the additional moment from the force about the new origin as well as rotation. Rotating the force vector alone is insufficient.
Verification: Define both origins and orientations and verify signs with a known lever-arm load.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
44Can a compensation matrix from another sensor be copied directly?
Not without verification. A matrix relates to particular channels, sensitivities, units, coordinates and calibration conditions, potentially varying by unit or revision.
Verification: Check serial identity, channel order and units and verify with independent loads.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
45What happens if N mm and N m are mixed?
They differ by a factor of one thousand. Mixed matrix, display or alarm units can create scale errors and incorrect overload decisions.
Verification: Mark units at every stage and test the chain with a known force and lever arm.
Limites d’application : Moments depend on reference origin; coordinate or unit errors alter results
46Can asynchronous axis sampling distort a dynamic resultant?
Yes. Components from different times do not represent one physical load state during fast changes and can distort resultant and compensated values.
Verification: Review synchronisation and channel delays and test timing with a common dynamic event.
Limites d’application : A compensation matrix must match the sensor and calibrated scope
47Can installation preload contribute to the measurement baseline?
It can. Flange, bolt and mating-surface stresses can enter the measuring structure; uneven or excessive preload can also affect error. Follow prescribed installation.
Verification: Record tightening sequence and torque and inspect raw baselines and remounting repeatability.
Limites d’application : Individual axis ratings do not prove every combined load is permissible
48Can a multiaxis sensor directly replace a robot collision-safety system?
Measuring force alone does not grant a safety function. Detection, control, actuation and fault response require the relevant risk assessment and evidence.
Verification: Have the safety owner define and validate the function using model-specific evidence.
Limites d’application : Check crosstalk definitions and the affected-axis full-scale reference
49Why can a heavily loaded axis matter more when measuring a small component?
Even modest full-scale crosstalk can be large relative to a small target component. Assess the actual load combination and error definition, not one percentage alone.
Verification: Assess maximum interfering and minimum target loads and test representative combinations.
Limites d’application : Moments depend on reference origin; coordinate or unit errors alter results
50Does a single-axis calibration establish six-axis combined accuracy?
It does not establish every combination. Assess sensitivities, crosstalk, linearity and mounting for the task and agree required combined-load evidence before acceptance.
Verification: Review calibrated axes, loads, directions and origins and identify combinations needing additional evidence.
Limites d’application : A compensation matrix must match the sensor and calibrated scope






