La version française n’est pas encore disponible. La version complète en anglais est présentée ci-dessous.
← Retour à la base de connaissancesMagnetic level indicators: 50 selection, application and maintenance questions
50 shared questions for Magnetic level indicators, covering selection, delivery, maintenance, common questions and evidence-based brand comparisons, with verification actions, limits and sources.
Conservez vos ressources techniques
Conservez un brouillon dans ce navigateur, puis importez-le manuellement dans votre bibliothèque après connexion.
La création de compte suit le processus habituel d’inscription et d’approbation. L’enregistrement de ressources n’ouvre pas automatiquement l’accès aux prix ni aux autres autorisations commerciales.
Shared guidance for this product family. Confirm the complete model and application; family-level guidance is not evidence of a particular product specification or approval.
Les principes généraux s’appliquent en complément de la validation du modèle. Les performances, homologations et alternatives dépendent du code de commande complet et des conditions réelles.
1What does the Magnetic level indicators principle establish, and what does it not establish?
A magnetic float in a communicating chamber drives an external display or detector; chamber communication and float suitability are essential. The principle explains signal generation; it does not establish the range, accuracy, materials or approvals of every model. Review the physical principle and the application limits separately.
Verification: Confirm "Actual minimum density" and assess "Blocked chamber connections invalidate representation" against the actual application. Record the complete model and document revision used.
Limites d’application : Blocked chamber connections invalidate representation
2Is the required result measured directly or calculated from other inputs?
Define the quantity to be accepted before selecting a display unit. A magnetic float in a communicating chamber drives an external display or detector; chamber communication and float suitability are essential. For "Actual minimum density", document output definitions and conversion or compensation conditions; displaying a value does not make it a direct measurement under all conditions.
Verification: Document the process objective, units and reference conditions. Check "Blocked chamber connections invalidate representation" and list separate inputs and acceptance methods for calculated results.
Limites d’application : Blocked chamber connections invalidate representation
3Which application information is needed before requesting a quotation?
Prepare at least these five input groups: Actual minimum density; Chamber and float temperature/pressure limits; Tap spacing and connection type; Viscosity, crystallization and deposits; Local indication and remote output. Mark unknown items for confirmation instead of substituting a normal point or an assumed default for all conditions.
Verification: Have process, instrumentation and purchasing roles agree the inputs, particularly "Chamber and float temperature/pressure limits". Resolve missing evidence before requesting application confirmation.
Limites d’application : Blocked chamber connections invalidate representation
4How should the range be chosen rather than copied from a nominal equipment rating?
Select the range from actual operating conditions and permissible error, then check extremes and installation constraints. For this family: Blocked chamber connections invalidate representation Nominal size or maximum equipment rating need not equal the useful measuring interval.
Verification: List minimum, normal, maximum and abnormal conditions. Check "Tap spacing and connection type" against the documented performance envelope.
Limites d’application : Blocked chamber connections invalidate representation
5Can a larger turndown claim or display rescaling fix poor low-end performance?
No. Low-end usability depends on the principle, available signal and error mechanisms. Check float and chamber pressure ratings separately More display digits, software scaling or a stated turndown cannot remove the original noise or installation error.
Verification: Obtain error definitions and limits at the actual low operating point. Use "Why can remote output disagree with the flap display → Display and transmitter paths can have different alignment or scaling errors. → Check upper/lower connections, venting and drains and record Viscosity, crystallization and deposits" to investigate bias instead of comparing only percent-of-full-scale values.
Limites d’application : Check float and chamber pressure ratings separately
6How should compatibility with the measured object and materials be established?
Use the complete measured object and actual exposure, rather than a material name alone. Viscosity, crystallization and deposits Avoid this family-specific misuse: Check float and chamber pressure ratings separately For non-wetted equipment, assess loads, target materials, signals and environmental compatibility instead.
Verification: List the actual contact or loading parts, seals, insulation and cleaning conditions. Have the responsible parties assess "A flap display does not inherently provide remote output".
Limites d’application : A flap display does not inherently provide remote output
7Why must abnormal and environmental conditions be listed separately from normal operation?
Normal measurement, temporary excursions and survival limits are different requirements. A flap display does not inherently provide remote output Start-up, cleaning, impacts, power interruption or abnormal environments can determine the necessary construction and protection.
Verification: Complete the worst-case conditions using "Local indication and remote output" and check "Interface and surface floats are not interchangeable". Separate continuous operating limits from temporary limits in the specification.
Limites d’application : Interface and surface floats are not interchangeable
8Can the published instrument accuracy be used as the installed system accuracy?
Not without assessment. Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment. The installed system also includes object, mounting, signal conversion, environmental and reference effects. Derived results additionally depend on the relevant input uncertainties.
Verification: Agree actual test points, installation conditions and the acceptance reference. Check "Check upper/lower connections, venting and drains" and document the system error or uncertainty basis.
Limites d’application : Check float and chamber pressure ratings separately
9How should response time, filtering and a stable display be balanced?
A stable indication and fast change detection involve a trade-off; filtering cannot repair incorrect installation or real process variation. Why can a float sink at high temperature → Reduced density or float deformation can compromise buoyancy. → Admit process pressure using the approved sequence and record Tap spacing and connection type Review the complete chain when fast response matters, not merely the display refresh rate.
Verification: Define the shortest relevant event and acceptable delay. Check "Check initial display state per the manual" and verify local output, display and receiving-system response using a controlled change.
Limites d’application : A flap display does not inherently provide remote output
10Which typical applications fit, and which require a specific engineering review?
Typical contexts include: Local tank indication; Process-vessel bypass level; Local versus remote comparison. Similar application labels do not establish equivalent conditions. In particular, assess "Interface and surface floats are not interchangeable" before relying on another installation as proof.
Verification: Select the nearest application and compare "Actual minimum density" and "Local indication and remote output". Obtain written confirmation where published application limits are exceeded.
Limites d’application : Interface and surface floats are not interchangeable
11How should a complete order code be defined without unspecified standard options?
Translate required functions, performance and physical interfaces into verifiable options. Actual minimum density; Chamber and float temperature/pressure limits. Options affecting suitability or supply scope should be explicit; terms such as standard or same as existing are insufficient.
Verification: Check the complete order code and selection sheet, especially "Tap spacing and connection type". List accessories, documents, calibration and delivery boundaries separately.
Limites d’application : Blocked chamber connections invalidate representation
12Which mounting and mechanical layout checks belong before ordering?
Check upper/lower connections, venting and drains The installation drawing must match the available space and process or mechanical structure. Access for fitting and servicing should be established before delivery.
Verification: Review a dimensioned, oriented site drawing. Use "Viscosity, crystallization and deposits" and "A flap display does not inherently provide remote output" to confirm connections, support, removal access and safe isolation.
Limites d’application : A flap display does not inherently provide remote output
13How can supporting infrastructure prevent an instrument that fits but cannot be used?
Ensure correct float orientation and travel Supporting conditions are part of the measurement system. Matching mechanical connections or signal names alone do not establish power, I/O, grounding or configuration compatibility.
Verification: Confirm "Check upper/lower connections, venting and drains" and "Verify output, failure state and receiving-system compatibility" against actual terminals, loops and receiving-system settings.
Limites d’application : Check float and chamber pressure ratings separately
14What is a sensible commissioning sequence before changing correction factors?
Check identity, installation and basic conditions first, then status, output and process response. Check initial display state per the manual Preserve original evidence for unexplained anomalies instead of concealing them with zero or factor adjustments.
Verification: Record factory settings and initial status. Investigate "Why can indication remain fixed while vessel level changes → Connections, float travel, integrity or magnetic coupling may be faulty. → Check initial display state per the manual and record Chamber and float temperature/pressure limits" step by step and change parameters only after identifying the cause and using an authorised procedure.
Limites d’application : A flap display does not inherently provide remote output
15Do identical output units or interface names establish connection compatibility?
No. Establish the represented object and reference conditions: Actual minimum density. Electrical equipment also needs loop type, scaling, failure behaviour and protocol details checked. Mechanical indicators require clear boundaries between indication and remote-transmission accessories.
Verification: Check the receiving-system variable against "Actual minimum density" and verify limits, units and abnormal states with the actual equipment. Record wiring, configuration and end-to-end checks against "Ensure correct float orientation and travel".
Limites d’application : Check float and chamber pressure ratings separately
16What should a calibration report contain to be relevant to the application?
Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment. The report should identify the delivered instrument, measured quantity, conditions, method and results. A general conformity certificate does not establish performance for every option or operating condition.
Verification: Check serial number, points, units, reference equipment and traceability. Compare "Chamber and float temperature/pressure limits" with the calibrated conditions and agree separate acceptance where they differ.
Limites d’application : Check float and chamber pressure ratings separately
17How can delivery and site acceptance use reproducible pass criteria?
Admit process pressure using the approved sequence Agree criteria before testing, including range, conditions, error definition and treatment of anomalies. Similar-looking readings or an unsuitable portable reference are not sufficient acceptance evidence.
Verification: Choose an appropriate reference using "Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment.". Record "Actual minimum density" and test conditions; investigate "Why can remote output disagree with the flap display → Display and transmitter paths can have different alignment or scaling errors. → Check upper/lower connections, venting and drains and record Viscosity, crystallization and deposits" before deciding on rework or recalibration.
Limites d’application : Check float and chamber pressure ratings separately
18What belongs in the quotation and agreement for a non-standard application?
Define why the standard supply is insufficient, the required verification and the responsible parties. Interface and surface floats are not interchangeable; Interface and surface floats are not interchangeable. Additional materials, construction, documents or tests depend on actual requirements.
Verification: Attach "Local indication and remote output" and the deviations. Obtain technical confirmation, additional charges, delivery conditions and acceptance methods; unknown requirements must not be assumed satisfied.
Limites d’application : Interface and surface floats are not interchangeable
19Where should investigation start when there is no valid reading or output?
Why can indication remain fixed while vessel level changes → Connections, float travel, integrity or magnetic coupling may be faulty. → Check initial display state per the manual and record Chamber and float temperature/pressure limits Distinguish the actual measured condition from device-state and readout or transmission faults. Do not immediately interpret missing data as sensor failure or a zero process value.
Verification: Under appropriate safe isolation, check "Ensure correct float orientation and travel". Compare local and receiving-end status and record when the fault appeared.
Limites d’application : Blocked chamber connections invalidate representation
20How can unstable readings be separated from genuine process variation?
Why can a float sink at high temperature → Reduced density or float deformation can compromise buoyancy. → Admit process pressure using the approved sequence and record Tap spacing and connection type Compare independent process evidence and raw signals before deciding whether installation, interference or equipment is responsible. More filtering can hide a real abnormal event.
Verification: Keep conditions comparable and record correlation with process or equipment events. Check "Check upper/lower connections, venting and drains" and change one factor at a time.
Limites d’application : A flap display does not inherently provide remote output
21How should a persistent high or low bias be investigated before changing factors?
Why can remote output disagree with the flap display → Display and transmitter paths can have different alignment or scaling errors. → Check upper/lower connections, venting and drains and record Viscosity, crystallization and deposits Systematic bias can arise from units, reference conditions, configuration, installation or the reference method. An unexplained correction can move one operating-point error to other conditions.
Verification: Align quantities, time bases and units first. Review the reference using "Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment." and confirm that "Chamber and float temperature/pressure limits" matches the configuration.
Limites d’application : Check float and chamber pressure ratings separately
22When is it appropriate to adjust a drifting zero or reference position?
Can insulation cover the whole instrument → No. Chamber temperature needs and electronic cooling/readability must be assessed separately. → Ensure correct float orientation and travel and record Local indication and remote output Establish a genuine, stable reference condition that meets the method requirements before considering adjustment. Some instruments have no freely adjustable site zero.
Verification: Check "Check initial display state per the manual" and "Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment." and retain before-and-after results. If the conditions cannot be established, remove external effects or use a qualified service.
Limites d’application : A flap display does not inherently provide remote output
23Which evidence helps diagnose intermittent alarms or failures?
Is a drain valve enough for crystallizing media → No. Heating, flushing, connection geometry and isolation must be evaluated. → Check initial display state per the manual and record Actual minimum density Intermittent faults need correlated event records, not just a normal display after recovery. Review alarm settings, communication status and actual process abnormalities separately.
Verification: Retain time, status codes, raw trends and the contemporaneous "Local indication and remote output". Compare "Ensure correct float orientation and travel" and "Check initial display state per the manual" to locate the failing link.
Limites d’application : Interface and surface floats are not interchangeable
24Why should repeated failures trigger a cause review instead of another identical replacement?
Repeated failures warrant a common-cause assessment. Interface and surface floats are not interchangeable If the external conditions remain unchanged, an identical construction or installation can experience the same failure again.
Verification: Summarise failed parts, service time and "Local indication and remote output". Check "Check upper/lower connections, venting and drains" and "A flap display does not inherently provide remote output" and establish corrective action before selecting repair or redesign.
Limites d’application : A flap display does not inherently provide remote output
25What distinguishes a site comparison, an adjustment and a calibration?
Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment. A comparison identifies differences; adjustment changes indication or parameters; calibration establishes a reference relationship under stated conditions. They are not interchangeable claims.
Verification: Define what performance and evidence must be established, using conditions relevant to "Chamber and float temperature/pressure limits". Retain original and adjusted results, references and responsibility records.
Limites d’application : Check float and chamber pressure ratings separately
26What must be checked again after replacing an instrument with the same basic model?
The same basic model does not establish identical options, software, calibration or accessories. Verify output, failure state and receiving-system compatibility; Confirm certificates, calibration and site change approval in writing. Previous system operation does not remove the need to verify a replacement.
Verification: Compare full order codes and revisions. Recheck critical points, failure states and interfaces using "Admit process pressure using the approved sequence" before handover.
Limites d’application : Interface and surface floats are not interchangeable
27Can every published maximum rating be achieved simultaneously?
No. Blocked chamber connections invalidate representation Maximum values usually relate to particular options and conditions. Temperature, load, pressure, materials, speed or signal conditions may constrain one another.
Verification: Review "Tap spacing and connection type" and "Local indication and remote output" together. Obtain confirmation for the combined conditions instead of assembling unrelated maxima from separate pages.
Limites d’application : Blocked chamber connections invalidate representation
28Does a missing signal or zero indication establish that the process value is zero?
No. Why can indication remain fixed while vessel level changes → Connections, float travel, integrity or magnetic coupling may be faulty. → Check initial display state per the manual and record Chamber and float temperature/pressure limits Indication, measurement and transmission have distinct possible failure modes. Also consider "Blocked chamber connections invalidate representation"; establish the actual process condition independently.
Verification: Check available status or failure indications and "Check upper/lower connections, venting and drains". For interlocks, follow the complete safety design rather than substituting a zero value for diagnosis.
Limites d’application : Blocked chamber connections invalidate representation
29Does one acceptable test point establish compliance across the working range?
No. Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment. Test points and directions must reflect the working interval, relevant error mechanisms and acceptance purpose. Multiple points, reversal or load tests depend on the method.
Verification: Define acceptance coverage against "Chamber and float temperature/pressure limits" and include critical operating points. Mark conditions outside the tests as unverified.
Limites d’application : Check float and chamber pressure ratings separately
30Are the largest range, highest accuracy class and most options always preferable?
Not necessarily. Options should serve the required objective; excess configuration can add cost, complexity or incompatibility. Check float and chamber pressure ratings separately; Check float and chamber pressure ratings separately. Establish essential requirements before comparing optional benefits.
Verification: Document the objective in "Actual minimum density" and why "Actual minimum density" is needed. Compare performance at actual working points rather than one headline metric.
Limites d’application : Check float and chamber pressure ratings separately
31Can every maintenance or recalibration interval simply be fixed at one year?
Not independently of risk and history. Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment. Intervals depend on applicable law or contract, conditions, drift, maintenance and failure evidence. Generic advice cannot replace mandatory requirements.
Verification: Use "Can insulation cover the whole instrument → No. Chamber temperature needs and electronic cooling/readability must be assessed separately. → Ensure correct float orientation and travel and record Local indication and remote output" and "Interface and surface floats are not interchangeable" to establish and review an initial plan. Retain the justification and approval for interval changes.
Limites d’application : Interface and surface floats are not interchangeable
32Can parts from different brands be mixed if their physical interfaces fit?
Physical fit is insufficient. Chamber and float temperature/pressure limits; Check upper/lower connections, venting and drains. Matching, materials, signals, calibration and approval boundaries can change with the combination.
Verification: Check the complete combination against "Ensure correct float orientation and travel" and obtain compatibility and acceptance evidence. An unconfirmed assembly must not be described as an approved system.
Limites d’application : Check float and chamber pressure ratings separately
33Does a hazardous-area, hygienic or safety label establish suitability for that duty?
Check the exact order configuration and valid documents rather than a series name. Confirm certificates, calibration and site change approval in writing "Interface and surface floats are not interchangeable" also matters; certificates, installation conditions and restrictions together define suitability.
Verification: Obtain certificate identifiers, covered models and special conditions and match them to the site. Similar names do not grant functional-safety or legal-metrology approval.
Limites d’application : Interface and surface floats are not interchangeable
34What evidence makes the final result acceptable, reproducible and sustainable?
Establish the chain from quantity definition and application limits through configuration, mounting, reference and operating status. Admit process pressure using the approved sequence A normal indication or brochure alone cannot demonstrate continuing system reliability.
Verification: Put "Actual minimum density", "Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment." and "Is a drain valve enough for crystallizing media → No. Heating, flushing, connection geometry and isolation must be evaluated. → Check initial display state per the manual and record Actual minimum density" into the technical, acceptance and maintenance records, with responsibility for reviewing excursions.
Limites d’application : Interface and surface floats are not interchangeable
35Can a DOWESTON model replace an installed instrument from another brand?
Possibly, after a full configuration and application comparison. A basic model name or matching output is not proof of equivalent replacement. Actual minimum density; Chamber and float temperature/pressure limits; Check upper/lower connections, venting and drains. Record conclusions on material differences.
Verification: Provide the original full order code, data sheet and site conditions. Check "Verify output, failure state and receiving-system compatibility" and "Confirm certificates, calibration and site change approval in writing" and obtain the necessary test and approval evidence before replacement.
Limites d’application : Interface and surface floats are not interchangeable
36Why choose DOWESTON, and what would make that reasoning evidence-based?
Base the decision on project compliance and documented delivery or service, rather than a slogan. "Actual minimum density" and "Blocked chamber connections invalidate representation" are important family-specific checks. Claim only benefits supported by the offered product documents and written commitments.
Verification: Request the proposed configuration matching "Actual minimum density", test evidence and delivery/service terms. Do not infer unverified performance, approvals, origin or service life.
Limites d’application : Blocked chamber connections invalidate representation
37How can competing brands be compared beyond a single accuracy or price figure?
Use identical conditions and supply boundaries. Actual minimum density; Check upper/lower connections, venting and drains; Confirm certificates, calibration and site change approval in writing. Quotes with different ranges, compensation, accessories or calibration scope can conceal material differences.
Verification: Compare "Chamber and float temperature/pressure limits", "Check float and chamber pressure ratings separately" and mounting/acceptance conditions. Separate demonstrated, unverified and inapplicable items; do not claim unproven overall superiority.
Limites d’application : Check float and chamber pressure ratings separately
38How should purchasing compare cost without overlooking additions or rework?
Compare complete supply and use-cycle costs, not only the main instrument price. Check upper/lower connections, venting and drains; Use a traceable reference appropriate to the principle, compare several application points and record conditions before and after adjustment. List the actual installation, supporting equipment, calibration, consumables, downtime and service requirements without assuming any brand is always cheaper.
Verification: Use the same "Actual minimum density" configuration and "Local indication and remote output" boundary. Separate equipment, accessories, commissioning and maintenance responsibilities and retain a traceable comparison.
Limites d’application : Interface and surface floats are not interchangeable
39Why does a bypass chamber represent vessel level?
Correct upper and lower communication establishes the corresponding fluid and pressure state. Verify Check upper/lower connections, venting and drains and record Actual minimum density. Blocked chamber connections invalidate representation. Confirm the exact model and record the result before changing settings.
Verification: Check upper/lower connections, venting and drains and record Actual minimum density
Limites d’application : Blocked chamber connections invalidate representation
40Is equal tap spacing sufficient for replacement?
No. Connection standard, seals, bore, float and drain arrangements also matter. Verify Ensure correct float orientation and travel and record Chamber and float temperature/pressure limits. Check float and chamber pressure ratings separately. Confirm the exact model and record the result before changing settings.
Verification: Ensure correct float orientation and travel and record Chamber and float temperature/pressure limits
Limites d’application : Check float and chamber pressure ratings separately
41Why assess minimum operating density?
The float must remain buoyant under the lowest actual process density. Verify Check initial display state per the manual and record Tap spacing and connection type. A flap display does not inherently provide remote output. Confirm the exact model and record the result before changing settings.
Verification: Check initial display state per the manual and record Tap spacing and connection type
Limites d’application : A flap display does not inherently provide remote output
42Can the float be installed in either direction?
Magnet geometry and mechanical limits can make orientation critical. Verify Admit process pressure using the approved sequence and record Viscosity, crystallization and deposits. Interface and surface floats are not interchangeable. Confirm the exact model and record the result before changing settings.
Verification: Admit process pressure using the approved sequence and record Viscosity, crystallization and deposits
Limites d’application : Interface and surface floats are not interchangeable
43Does an all-white display prove an empty vessel?
No. It may be an initial, stuck or undriven display state. Verify Check upper/lower connections, venting and drains and record Local indication and remote output. Blocked chamber connections invalidate representation. Confirm the exact model and record the result before changing settings.
Verification: Check upper/lower connections, venting and drains and record Local indication and remote output
Limites d’application : Blocked chamber connections invalidate representation
44Why avoid rapid pressurization?
Rapid admission can cause float impact or transient loading. Verify Ensure correct float orientation and travel and record Actual minimum density. Check float and chamber pressure ratings separately. Confirm the exact model and record the result before changing settings.
Verification: Ensure correct float orientation and travel and record Actual minimum density
Limites d’application : Check float and chamber pressure ratings separately
45Why can indication remain fixed while vessel level changes?
Connections, float travel, integrity or magnetic coupling may be faulty. Verify Check initial display state per the manual and record Chamber and float temperature/pressure limits. A flap display does not inherently provide remote output. Confirm the exact model and record the result before changing settings.
Verification: Check initial display state per the manual and record Chamber and float temperature/pressure limits
Limites d’application : A flap display does not inherently provide remote output
46Why can a float sink at high temperature?
Reduced density or float deformation can compromise buoyancy. Verify Admit process pressure using the approved sequence and record Tap spacing and connection type. Interface and surface floats are not interchangeable. Confirm the exact model and record the result before changing settings.
Verification: Admit process pressure using the approved sequence and record Tap spacing and connection type
Limites d’application : Interface and surface floats are not interchangeable
47Why can remote output disagree with the flap display?
Display and transmitter paths can have different alignment or scaling errors. Verify Check upper/lower connections, venting and drains and record Viscosity, crystallization and deposits. Blocked chamber connections invalidate representation. Confirm the exact model and record the result before changing settings.
Verification: Check upper/lower connections, venting and drains and record Viscosity, crystallization and deposits
Limites d’application : Blocked chamber connections invalidate representation
48Can insulation cover the whole instrument?
No. Chamber temperature needs and electronic cooling/readability must be assessed separately. Verify Ensure correct float orientation and travel and record Local indication and remote output. Check float and chamber pressure ratings separately. Confirm the exact model and record the result before changing settings.
Verification: Ensure correct float orientation and travel and record Local indication and remote output
Limites d’application : Check float and chamber pressure ratings separately
49Is a drain valve enough for crystallizing media?
No. Heating, flushing, connection geometry and isolation must be evaluated. Verify Check initial display state per the manual and record Actual minimum density. A flap display does not inherently provide remote output. Confirm the exact model and record the result before changing settings.
Verification: Check initial display state per the manual and record Actual minimum density
Limites d’application : A flap display does not inherently provide remote output
50How should a replacement be accepted?
Compare chamber/float data and test local and remote indication under controlled level conditions. Verify Admit process pressure using the approved sequence and record Chamber and float temperature/pressure limits. Interface and surface floats are not interchangeable. Confirm the exact model and record the result before changing settings.
Verification: Admit process pressure using the approved sequence and record Chamber and float temperature/pressure limits
Limites d’application : Interface and surface floats are not interchangeable






