Parta de su proceso y encuentre el instrumento adecuado
Aprenda los principios de medición, compruebe las condiciones del proceso, configure productos y guarde recursos técnicos.

La edición en español aún no está disponible. A continuación se muestra la versión completa en inglés.

← Volver a la base de conocimientos
Principios de medición5 min de lecturaIdioma del contenido: inglésActualizado 23 de octubre de 2026

Gauge or absolute pressure sensors: what is the difference and which should you choose?

Confirm the zero reference before comparing range and output. Identical units do not mean the measurements represent the same pressure.

The reference pressure is the difference

Gauge pressure is the difference between the absolute process pressure and local atmospheric pressure. Absolute pressure is measured relative to a perfect vacuum. With the process port open to the local atmosphere, a vented gauge sensor should read close to zero, while an absolute sensor should read close to the local atmospheric pressure. Even when both readings are expressed in bar or kPa, they cannot be compared directly. Specify the pressure type on the order.

Choose according to the control objective

For pneumatic or hydraulic equipment where pressure relative to the surroundings matters, start by evaluating a gauge-pressure sensor. Vacuum processes and gas-state calculations that require absolute pressure should use an absolute sensor or a conversion method that has been justified. Adding a fixed atmospheric pressure in the program does not make the results equivalent at different elevations or under different weather conditions.

Keep the reference consistent when replacing an instrument

Give Doweston the old nameplate details, the control objective, the installation site and the PLC range settings. In particular, clarify whether “zero pressure” means vented to atmosphere or evacuated. During commissioning, compare the reference instrument and the sensor under test using the same pressure reference before assessing zero offset. Do not adjust zero solely because the readout is above or below zero.

Field checks: define the pressure reference first

  • State gauge or absolute pressure in the process documentation and verify that the original instrument, the reference instrument and the program use the same reference.
  • Record the expected readings when vented, during normal operation and under vacuum. Check what “zero pressure” means in each case.
  • For gas-state calculations, check whether the formula requires absolute pressure before using a gauge reading.
  • For equipment used at several sites, document whether local elevation and atmospheric changes affect the required result. Check other sources of error, such as temperature, separately.

Example: a difference of one atmosphere may be normal

Suppose a customer replaces a gauge sensor with an absolute sensor of the same numerical range. The new sensor does not read zero with the port vented. If the local atmospheric pressure is assumed to be about one hundred kilopascals, an absolute reading close to that value reflects the different reference. It does not indicate a damaged zero point. This value is for illustration only and must not be treated as a universal conversion constant.

First establish the process objective: does the system control the remaining absolute pressure in the vessel or the pressure above the surroundings? Then confirm the sensor version and the program against that reference. The sales team should request the full original nameplate and a written description of the expected reading under one known condition. This is usually more effective at identifying an incorrect reference than asking for the desired pressure value alone.

Aplique estos conocimientos a la selección

Clientes y compras

Compruebe las condiciones reales y los criterios de aceptación con el artículo y la documentación del producto. Envíe las cuestiones pendientes al servicio técnico de DOWESTON.

Equipos comerciales y técnicos

Anote las preguntas del cliente y sus condiciones de servicio. Compruebe el rango, las conexiones y las condiciones de uso en la documentación del modelo seleccionado.

Convierta las preguntas en criterios de selección

Reúna los datos del fluido, las conexiones, la instalación y los criterios de aceptación del producto, y revise los puntos pendientes con DOWESTON.

Elegir un producto y preparar las condiciones de servicio
¿Aún no ha iniciado sesión en DOWESTON?
Cree una cuenta para empezar a seleccionar, consultar precios y recibir nuestro boletín con las últimas novedades.
Productos y servicios
Doweston ofrece una gama de servicios para ayudarle a construir rápidamente sistemas de instrumentación automatizada. Contáctenos para más información.
¿Listo para comprar?
Si desea comprar productos,
solo contacte con nuestro equipo de ventas — estaremos encantados de ayudarle.
Quiénes somos
DOWESTON ofrece soluciones de medición precisas y fiables
¡productos de detección y sensores fiables para todos los sectores!
¿Necesita servicio?
Si desea optimizar la producción y reducir los parones,
podemos ayudarle a elegir las piezas de repuesto y los productos de servicio adecuados.
Editar configuración

Nombre del producto

Marca

Modelo

Confirmar cambios Cancelar
Su pantalla puede ser demasiado pequeña para una mejor experiencia.
Le recomendamos continuar en un ordenador,
¡o contacte de inmediato con un ingeniero para su selección!
¿Servicio más rápido? Llame al
400-888-4436
Seguir navegando